Glutamate Decarboxylase Assay
Method: The reaction velocity is measured by a conventional Warburg manometric technique where glutamate is converted to aminobutyrate. One unit releases one micromole of CO2 per minute at 37°C and pH 5.0 under the specified conditions.
Reagents
Enzyme Immediately prior to use, dilute the enzyme with reagent grade water to a concentration of 0.5 - 3.0 units/ml. Procedure Into the main well of the Warburg flask pipette the following:
Pipette 0.5 ml of 0.05 M L-glutamate into side arm. Include one flask with no enzyme as a blank and one flask with 3.0 ml water as a thermal barometer. Assemble flasks and incubate for 10 - 15 minutes to achieve temperature equilibration. Tip in substrate and measure CO2 evolution at 2 - 3 minute intervals for 30 minutes. Calculation |